Neonatal equipment holds a very small patient inside a very narrow tolerance. A warmer two degrees out or a phototherapy unit at half its rated irradiance is not a minor fault. We service labour rooms, SNCUs and NICUs across the district health system and private hospitals in India.

FIG. 07 / GYNAECOLOGY & NEONATAL
A bedside multi-parameter patient monitor showing live traces at a made-up bed.

Equipment covered

Supplied, installed, maintained and repaired. Anything that measures or delivers a quantity is calibrated and documented at every visit.

  • 01Infant radiant warmer
  • 02Neonatal incubator
  • 03Phototherapy unit
  • 04Transcutaneous jaundice meter
  • 05Fetal monitor
  • 06Fetal doppler
  • 07Gynaecology OT spot light
  • 08Delivery table
  • 09Paediatric suction machine
  • 10Nebuliser
  • 11Infant weighing scale

How this department is supported

Maintenance contracts

AMC and CMC cover with scheduled preventive visits and calibration at documented intervals.

Contract Scope

Breakdown repair

On-site diagnosis, with workshop support at Dudheswar for component-level restoration.

Repair Protocols

Supply & installation

Equipment supplied, installed, commissioned and calibrated, then maintained by the same engineers.

Request Quotation
Department FAQ

Questions we get about gynaecology & neonatal

Not answered here? Call +91 98250 70790 and speak to someone who works on the equipment.

01

How do you check whether a phototherapy unit still works?

With a radiometer, at the surface where the baby lies. Irradiance falls steadily with lamp hours long before any lamp fails, and a unit that looks bright can be well under the effective threshold. There is no way to judge this by eye, which is why units that appear fine are often not treating.

02

Our radiant warmer skin probe reads differently from the thermometer. Which is right?

Test both against a reference. Skin probes drift and are also the most frequently damaged item on a warmer. The cable fails at the strain relief and the reading shifts before it fails outright. Probe replacement and controller recalibration are routine work and should not wait for the next scheduled visit.

03

Do you service equipment in government SNCUs and CHCs?

Yes. A substantial part of our work over the last two decades has been in the district health system across Gujarat, in civil and general hospitals, community health centres and urban health centres, including maintenance carried out under the ORET programme and to National Health Mission requirements.

Next department

Ophthalmology, ENT and dental equipment

Discuss your equipment
requirements.

Contract quotations follow a precise site inspection and equipment audit, not a phone call. Contact us to schedule an assessment.

Head Office

11 Vrundavan Park, Sec 22
Gandhinagar 382021
Gujarat, India

Workshop

A-207, Sumel Safal-6
Dudheswar, Ahmedabad 380004
Gujarat, India