Physiotherapy equipment
Physiotherapy equipment delivers a dose, and almost none of it has any way of telling the therapist that the dose has changed. Output drifts, treatment stops working, and the machine gives no sign at all. We service and calibrate physiotherapy equipment across hospital departments and private clinics in India.

Equipment covered
Supplied, installed, maintained and repaired. Anything that measures or delivers a quantity is calibrated and documented at every visit.
- 01Continuous passive motion (CPM) machine
- 02Electrical muscle stimulator (EMS)
- 03Interferential therapy (IFT) unit
- 04Laser therapy unit
- 05Traction machine
- 06Therapeutic ultrasound unit
- 07Shortwave diathermy
- 08Infrared lamp
How this department is supported
Maintenance contracts
AMC and CMC cover with scheduled preventive visits and calibration at documented intervals.
Contract ScopeBreakdown repair
On-site diagnosis, with workshop support at Dudheswar for component-level restoration.
Repair ProtocolsSupply & installation
Equipment supplied, installed, commissioned and calibrated, then maintained by the same engineers.
Request QuotationQuestions we get about physiotherapy
Not answered here? Call +91 98250 70790 and speak to someone who works on the equipment.
01Our ultrasound therapy head does not seem to be working, but the machine shows output.
Test it by placing the head in a shallow dish of water at full output. A working head disturbs the surface visibly. If the water is still, the transducer crystal has delaminated from the faceplate, which is the usual failure. The display reads what the machine is sending, not what the head is emitting.
02How do you calibrate an IFT or EMS unit?
Output current and frequency are measured across the working range under a defined load, at each channel and each mode, and compared against the machine`s stated specification. Where output has drifted it is adjusted and remeasured. Both sets of readings appear on the certificate.
03How often should physiotherapy equipment be checked?
Annually for output-based devices, half-yearly for anything in continuous use in a busy department, and immediately when a therapist reports that treatments have stopped producing the response they used to. That report is a genuine fault indication and it is worth acting on.
Gynaecology and neonatal equipment
Discuss your equipment
requirements.
Contract quotations follow a precise site inspection and equipment audit, not a phone call. Contact us to schedule an assessment.
Head Office
11 Vrundavan Park, Sec 22Gandhinagar 382021
Gujarat, India
Workshop
A-207, Sumel Safal-6Dudheswar, Ahmedabad 380004
Gujarat, India