Ophthalmology, ENT and dental equipment
Three departments that share a common problem: precision optics and fine mechanisms mounted on furniture that gets adjusted a hundred times a day. The optics are what make the equipment worth maintaining, and the mechanisms are what usually fail.

Equipment covered
Supplied, installed, maintained and repaired. Anything that measures or delivers a quantity is calibrated and documented at every visit.
- 01Table-mounted ophthalmic unit
- 02Advanced ophthalmic refraction unit
- 03Slit lamp
- 04A-scan biometer
- 05Ophthalmic operating microscope
- 06ENT examination workstation
- 07Dental chair and unit
- 08Dental handpiece
- 09Dental X-ray
How this department is supported
Maintenance contracts
AMC and CMC cover with scheduled preventive visits and calibration at documented intervals.
Contract ScopeBreakdown repair
On-site diagnosis, with workshop support at Dudheswar for component-level restoration.
Repair ProtocolsSupply & installation
Equipment supplied, installed, commissioned and calibrated, then maintained by the same engineers.
Request QuotationQuestions we get about ophthalmology, ent & dental
Not answered here? Call +91 98250 70790 and speak to someone who works on the equipment.
01There is haze inside our slit lamp optics. Can that be cleaned?
Usually yes. Internal haze in Gujarat is almost always fungal growth on element surfaces, driven by monsoon humidity, and it is treatable if it is caught in the first season or two. Left for years it etches the coating permanently. This is one of the strongest arguments for a post-monsoon service visit in an eye department.
02Our dental chair has started sinking under the patient.
Hydraulic seal or check valve on a hydraulic chair, actuator or control board on an electric one. Both are repairable. The cylinder is stripped, resealed and load-tested before it goes back. Chairs are replaced for this far more often than they need to be.
03Do you work on ophthalmic operating microscopes?
Yes: illumination, focus and zoom drives, coaxial path, optical cleaning and stand balance. Where a fault sits inside a sealed assembly on a current model we say so and route it to the manufacturer rather than opening something we cannot correctly recollimate.