Sterilisation equipment

An autoclave is a pressure vessel that a hospital relies on for sterility assurance, and both of those facts have consequences. We service, calibrate and supply sterilisation equipment for CSSD, theatre and laboratory across India.

FIG. 09 / STERILISATION
An open autoclave chamber loaded with wrapped sterile packs.
A gowned hospital porter pushing a stainless steel trolley down a long, brightly lit ward corridor.

Equipment covered

Supplied, installed, maintained and repaired. Anything that measures or delivers a quantity is calibrated and documented at every visit.

  • 01Horizontal autoclave
  • 02Vertical autoclave
  • 03Table-top autoclave
  • 04ETO steriliser
  • 05Hot air oven
  • 06UV sterilisation chamber
  • 07Fogger
  • 08Instrument washer

How this department is supported

Maintenance contracts

AMC and CMC cover with scheduled preventive visits and calibration at documented intervals.

Contract Scope

Breakdown repair

On-site diagnosis, with workshop support at Dudheswar for component-level restoration.

Repair Protocols

Supply & installation

Equipment supplied, installed, commissioned and calibrated, then maintained by the same engineers.

Request Quotation
Department FAQ

Questions we get about sterilisation

Not answered here? Call +91 98250 70790 and speak to someone who works on the equipment.

01

Our autoclave reaches pressure but loads come out wet. What causes that?

Wet loads are a drying-phase or steam-quality problem, not a pressure problem. The usual causes are a failed steam trap, a weak vacuum pump on a pre-vacuum machine, an overloaded chamber, or feed water carrying too much scale. It is worth fixing properly, because a wet pack is not a sterile pack once it is handled.

02

How often should an autoclave be calibrated?

Chamber temperature, pressure and cycle timing every six months, and after any repair affecting the heating, control or steam path. The safety valve is tested at the same interval. The gauge on the front tells you what the sensor believes; only an independent reference tells you what the chamber is actually doing.

03

Is scale really a problem here?

Yes, and it is the single largest cause of autoclave failure in Gujarat. Hard feed water deposits scale on elements and inside the chamber and boiler. Scaled elements overheat and fail, scaled sensors read late, cycles lengthen and steam quality drops. Descaling on a schedule, and treating the feed water, costs a fraction of the elements it saves.

Next department

Hospital furniture

Discuss your equipment
requirements.

Contract quotations follow a precise site inspection and equipment audit, not a phone call. Contact us to schedule an assessment.

Head Office

11 Vrundavan Park, Sec 22
Gandhinagar 382021
Gujarat, India

Workshop

A-207, Sumel Safal-6
Dudheswar, Ahmedabad 380004
Gujarat, India