Orthopaedic and neurosurgical power tools
Surgical power tools are the most abused equipment in a hospital: autoclaved every cycle, dropped occasionally, and expected to run at full torque afterwards. Handpiece overhaul is bench work, and we do it at our Ahmedabad workshop rather than sending it out of the state.

Equipment covered
Supplied, installed, maintained and repaired. Anything that measures or delivers a quantity is calibrated and documented at every visit.
- 01Drill handpiece
- 02Saw handpiece
- 03Reamer handpiece
- 04Motor unit
- 05Shaver and RF handpiece
- 06Server console
- 07Multiple power tool sets
- 08Neurosurgical power tools
- 09Orthopaedic attachments and couplings
- 10Neurosurgery retractor sets
How this department is supported
Maintenance contracts
AMC and CMC cover with scheduled preventive visits and calibration at documented intervals.
Contract ScopeBreakdown repair
On-site diagnosis, with workshop support at Dudheswar for component-level restoration.
Repair ProtocolsSupply & installation
Equipment supplied, installed, commissioned and calibrated, then maintained by the same engineers.
Request QuotationQuestions we get about orthopaedic
Not answered here? Call +91 98250 70790 and speak to someone who works on the equipment.
01Our drill handpiece has lost torque and runs hot. What causes that?
Almost always bearings. Repeated autoclaving drives out lubricant and admits moisture, the bearings roughen, friction rises, torque falls and heat goes up. Left alone it takes the rotor with it. Caught early it is a bearing set and a reseal; caught late it is a motor.
02Can a handpiece be repaired or does it have to be replaced?
Most can be repaired. Bearings, seals, collets, chucks, couplings, cables and gear sets are all serviceable. Replacement becomes the sensible answer when the rotor or stator is damaged on a discontinued model, and we tell you which situation you are in before doing any work.
03How often should surgical power tools be serviced?
Every six months for a set in regular theatre use, and immediately after any drop or any cycle where the tool was autoclaved wet. Sets used only occasionally still need annual attention, because dried lubricant and seized couplings come from standing still as much as from use.