Laboratory equipment
A laboratory instrument that has drifted still produces a result, and that is the problem. We maintain and calibrate analysers, centrifuges, refrigeration and safety cabinets for hospital and diagnostic laboratories across India.

Equipment covered
Supplied, installed, maintained and repaired. Anything that measures or delivers a quantity is calibrated and documented at every visit.
- 01Semi-automatic analyser
- 02ELISA reader and washer
- 03Centrifuge
- 04Blood bank refrigerator
- 05Biosafety cabinet
- 06pH meter
- 07Laboratory microscope
- 08Laboratory incubator
- 09Hot air oven
How this department is supported
Maintenance contracts
AMC and CMC cover with scheduled preventive visits and calibration at documented intervals.
Contract ScopeBreakdown repair
On-site diagnosis, with workshop support at Dudheswar for component-level restoration.
Repair ProtocolsSupply & installation
Equipment supplied, installed, commissioned and calibrated, then maintained by the same engineers.
Request QuotationQuestions we get about laboratory
Not answered here? Call +91 98250 70790 and speak to someone who works on the equipment.
01Our analyser readings have drifted against controls. Where do we start?
Usually the lamp, the flow path or the pipettor. Optical sources lose output with running hours long before they fail, so the reading falls off gradually. We measure lamp output, check the flow path for blockage and air, verify pipettor volume, then recalibrate. Replacing reagent lots first is the common wasted step.
02Our blood bank refrigerator alarms at night but reads normal in the morning. Why?
That pattern is a door gasket, a failing defrost cycle or a probe reading a different point from where the samples sit. It is found by logging temperature at several positions over a full cycle rather than by looking at the display. A refrigerator that alarms intermittently is telling you something real.
03How often should a biosafety cabinet be tested?
Airflow velocity and containment every six months, and after any filter change or relocation. HEPA filters load progressively; the blower compensates until it cannot, and the cabinet fails containment before anything visible happens. UV lamp output should be measured rather than assumed from whether the lamp lights.
Physiotherapy equipment
Discuss your equipment
requirements.
Contract quotations follow a precise site inspection and equipment audit, not a phone call. Contact us to schedule an assessment.